Typically error codes issues will have to come back to us for repair.

The only exceptions are error 06, 09 &10 these can be cleared by doing a factory clear function.

Press and hold the (+) and (-) buttons then momentarily press the On/Standby button and turn the unit on.

Keep the (+) and (-) buttons held until the 7 segment displays are lit, then release the (+) and (-) buttons, then immediately press and hold the Alarm Silence button, prior to the 'REV' being displayed.

Verify 'dnE CLR' is output on the seven segment displays.

Release the Alarm Silence button.

Si se displaya error 6 en la pantalla. La pantalla de SpO2 contiene "Err" en ámbar, y la pantalla de ritmo cardíaco contiene "E06" en ámbar, se trata de:

• Flash externo

• Alarma de memoria

• Memoria no Funcional

Generalmente las incidencias de códigos de error tendrá que volver a nosotros para su reparación.

Las únicas excepciones son un error, 06, 09 y 10 de estos se puede borrar por hacer una fábrica de función clara.

Mantenga pulsado los botones (+) y (-), luego presione momentáneamente el botón de espera y encender la unidad.

Mantenga pulsado los botones (+) Y (-), y el de encendido hasta que se celebró la muestra de 7 segmentos, a continuación, liberar a los (+) y (-) botones, inmediatamente después mantenga pulsada la Botón de silenciar la alarma, antes de la 'que se muestra REV.

Verificar 'dnE CLR' se emite en las pantallas de siete segmentos.

Suelte el Botón de silenciar la alarma.

...